Digital pH Meter, Bench Mounted

Digital pH Meter, Bench Mounted

Digital pH Meter, Bench Mounted
ranges -2 to 16 pH, ±399.9 mV; ±2000 mV -9.9 to 120°C to BS 1377:3
CodeSL063
British StandardBS1377-3
Approx. Weight (kg)1.1
Tariff901580 19 0
Click to save page as PDFSave Page to PDF
or
Call Impact +44 (0)1294 602626

Advanced microprocessor based pH meter.In addition to pH it also measure mV for Ion Selective Electrode use and ORP/Redox measurements.

    5 memorised buffers for calibration: pH 4.01, 6.86, 7.01, 9.18 and10.01
  • Automatic temperature compensation
  • Large display shows pH or mV and temperature values simultaneously

Supplied complete with a double junction glass pH electrode, stainless steel temperature probe and calibration solutions it is ready for use straight out of the box.

Temperature compensation and pH buffer recognition and calibration are automatic and the meter features a 2level LCD display to provide a simultaneous readout of pH (or mV) alongside the solution temperature. During calibration user guide graphic symbols are shown on the left-hand side of the display making the meter extremely easy to use. The main instructions for use and calibration all fit neatly onto the front cover of the meter.

Specification:

Range: pH-2.00 to 16.00 pH
Range: mV±399.9 mV; ±2000 mV

Resolution: Temperature

-9.9 to 120.0 °C
Resolution: pH0.01 pH
Resolution: mV0.1 mV; 1 mV

Resolution: Temperature

0.1 °C
Accuracy @ 20°C : pH±0.01 pH
Accuracy @ 20°C : mV±0.2 mV; ±1 mV

Accuracy @ 20°C : Temperature

±0.4 °C
pH Calibrationautomatic, 1 or 2 point with 5 memorised buffer values ( pH 4.01, 6.86, 7.01, 9.18 and 10.01)
Temperature Compensationautomatic with supplied probe or manual from 9.9 to 120 °C
pH Electrode

glass body, double junction, refillable, 1m cable BNC connector

Temperature Probesupplied
Input Impedance1012Ohm
Power Supply12Vdc adaptor supplied
Environment0 to 50 °C, maximum 95% RH
Dimensions240mm x 182mm x 74mm
Weight1.1kg